4 5
2.1 - EXTENSION TEST CABLE
(RPI PART #SCT026)
When testing, troubleshooting or repairing the STATIM au-
toclave, install the RPI Extension Test Cable. By attaching
the Extension Test Cable to the cover and main PC board
as instructed below, you will be able to extend the dis-
tance between the autoclave and the cover thus allowing
you to place the cover in a horizontal (not vertical) position
for easier viewing of the LCD display and easier access to
the keypad.
INSTRUCTIONS
STEP #1 - Disconnect the power to the autoclave. Remove
the cover. Disconnect the two cables from the main PC
board - one cable from the LCD display board and the
other from the keypad. Leave the cover in the horizontal
position (see Figure 2.1A).
STEP #2 - Connect the Extension Test Cable mating con-
nector to the LCD display board cable.
STEP #3 - If the OEM keypad is installed, connect the Ex-
tension Test Cable mating connector to the keypad cable
connector (see Figure 2.1B).
If the RPI keypad is installed, connect the Extension Test
Cable mating connector to the RPI keypad cable connec-
tor - making sure that both connectors have the “This Side
Up” label facing in the upward direction (see Figure 2.1C).
(Note: The RPI keypad cable connector does not have an
alignment key, so it is very important to connect as per in-
structions and Figure 2.1C).
STEP #4 - Connect the Extension Test Cable mating con-
nectors to the main PC board - making sure that the
Extension Test Cable connector has the “This Side Up”
label facing in the upward direction (see Figure 2.1A).
(Caution: If the Extension Test Cable is connected incor-
rectly, severe damage may occur to the autoclave.)
THIS
UP
SIDE
THIS
UP
SIDE
THIS
UP
SIDE
SIDE
UP
THIS
SIDE
UP
THIS
FIGURE 2.1C
FIGURE 2.1B
FIGURE 2.1A
2.2 - START SWITCH
(RPI PART #SCS021)
When testing the pump or decontaminating the steam
generator/boiler, it is suggested by the OEM to activate the
pump by shorting Pins #18 and #16 together. However,
the RPI Start Switch offers an easier and safer method. By
connecting the RPI Start Switch to the main PC board con-
nector (P1), you control the activation of the pump via a
hand-held switch. In addition, the RPI Start Switch has a
“hold feature” (i.e. running lock) for longer test periods.
INSTRUCTIONS
STEP #1 - Connect the RPI Start Switch connector to Pins
#18 and #16 on the “20 Pin Connector” (P1) located on the
main PC board. (See Figure 2.2 and note the following: To
guide the Start Switch connector to Pins #18 and #16,
start at the pin labeled “20”, count down one pin, which
is Pin #18, then count down one more pin, which is Pin
#16. Gently guide the Start Switch connector to both Pins
#18 and #16, leaving all other pins exposed.) (Caution: If
the Start Switch is connected to any other pins except #18
and #16, severe damage may occur to the autoclave.) See
“Device Activation Chart”, page 47.
STEP #2 - Once the Start Switch is connected, the next
step is to connect the Pump Tester Bottle (see 2.3 “Pump
Tester Bottle” Instructions, page 6). When testing the
pump, press and hold the “momentary” button for
approximately 25 seconds. When decontaminating the
steam generator/boiler, (see Caution note below), press
and slide the “momentary” button into the “hold” (i.e. run-
ning lock) position for approximately 3 minutes. (Caution:
When decontaminating the steam generator/boiler, the
reservoir must be full so that the pump does not run dry
during the process thus causing severe damage to the au-
toclave.) See page 39 - Figure 9.2 for complete steam
generator/boiler cleaning procedure.
FIGURE 2.2
TIP - To guide the Start Switch
Connector to Pins #18 and
#16, start at the pin labeled
“20”, count down one pin,
which is Pin #18, then count
down one more pin, which is
Pin #16. Gently guide the Start
Switch Connector to both Pin
#18 and #16, leaving all other
pins exposed.
LCD Display Board
Connector
Keypad Connector
Label
“THIS SIDE UP”
LCD Display
Board Connector
Extension
Test Cable
OEM
Keypad
Cable Connector
(with alignment key)
Press and hold for
Momentary feature
LCD Display
Board Connector
Extension
Test Cable
RPI
Keypad
Cable Connector
(without alignment key)
Start Switch
(RPI Part #SCS021)
Press and slide for the
“Hold” feature (Running Lock)